Fabius
®
Fault-Cause-Remedy
6-4
Cop
yr
ight
2000 Dr
aeger
M
edical,
In
c.
sec6_
fau
ltcausr
em
.fm
Paw HIGH
The upper alarm limit
(Paw) has been exceeded.
The hose is kinked.
The upper alarm limit
(Paw) has been set too
high.
Check the upper alarm limit (Paw)
setting.
Check the hose system.
Correct the upper alarm limit
(Pmax) setting.
Paw NEGATIVE
Insufficient fresh-gas
supply.
Increase the fresh-gas flow.
POWER SUPPLY
INOP
Power supply is fault.
Inform DrägerService.
PRESSURE INOP
Faulty pressure sensor.
Inform DrägerService.
PRESSURE LIMIT
Set tidal volume has not
been delivered completely.
Increase Pmax value.
Decrease V
T
value.
SET APL TO IPPV
The APL valve is switched
to MAN in the IPPV mode.
Switch the APL valve to
SPONT/IPPV.
SETTINGS LOST
The set parameters or
calibration values have
been lost.
Check the set parameters or
calibration values and correct, if
necessary.
VENTILATOR INOP
Internal machine fault.
IPPV mode is not
operational.
Switch Fabius off and on.
Inform DrägerService.
VOL ALARM OFF
The lower alarm limit for
MV is off in the IPPV
mode.
Set the appropriate lower alarm
limit for MV.
Fault
Cause
Remedy
Содержание Fabius 4116029
Страница 2: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 6: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 8: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 68: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 234: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 318: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 324: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 326: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...