Intermittent Positive Pressure Ventilation (IPPV): General
A prerequisite for IPPV is that the patient is supplied with a sufficient amount of fresh
gas. The APL valve selector must be set to the "SPONT/IPPV" position.
If the APL valve selector is not set to the "SPONT/IPPV" position, the pressure in the
breathing system will increase.
The safety valve of the patient system makes sure that no pressures greater than
80 hPa (mbar) build up in the system. The pressure limit (Pmax) can be adjusted on
the control box.