Fabius
®
Repair Instructions
3-153
Cop
yr
ight
2000 Dr
aeger
M
edical,
In
c.
sec3_r
epai
rinstruc_c.fm
• Replace pressure sensor with a new one.
• Reconnect pressure sensor connector to Pneumatics Control PCB.
• Fasten all electrical cables together using a new cable tie.
• Cut off any protruding length of cable tie.
• Mount pneumatics to Fabius.
• Mount cover plate of pneumatics to Fabius.
• Measure equivalent leakage current according to Test Certificate.
• Measure protective conductor resistance according to Test Certificate.
• Perform a function test according to Test Certificate.