4-11
4.4.1
Sealing rings M 07152 (2x)
O2
N2O
5.
Cylinder pressure reducer
O2, serial no.
5.1
N2O, serial no.
5.2
Replace diaphragms 2312600 every 6
years
Date of next replacement: _________
5.2.1
Annual output test
Next test:
Detach cylinder connection from Fabius
and connect to measuring tube 7900718.
Open cylinder slightly.
With supply pressure 10 bar > 60 L/min.
O
2
N
2
O
5.3
Sealing ring M 05045
With pin index pressure reducer
O2
N2O
5.4
2nd stage blow-off valves
Check once a year.
Connect test pressure reducer e.g. with
DIN pressure hose to output o pressure
reducer
Opening pressure 7.0
+ 3.0
bar
O2
N2O
5.5
Static test every 6 months
Test instructions:
Connect output to pressure gauge
(15 bar) test equipment part no. 2M 5223.
C
C
*
C
O
O
C
C
O
O
Содержание Fabius 4116029
Страница 2: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 6: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 8: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 68: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 234: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 318: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 324: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 326: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...