Fabius
®
Repair Instructions
3-36
Cop
yr
ight
2000 Dr
aeger
M
edical,
In
c.
sec3_
repa
ir
instr
uc_c.
fm
Fig. 20:
Adjusting voltage range on power transformer
• Connect power transformer cable connector to connection
1
of
Power Supply PCB.
Fig. 21:
Power Supply PCB; connecting power transformer cable connector
Check that new power transformer is wired according to
country-specific requirements. Adjust voltage range if
necessary (see Figure "
Adjusting voltage range on power
transformer
".
Make sure to use the correct fuses for 110 V or 230 V supply:
100 V = 2 AT, 230 V = 1 AT