Fabius
®
Repair Instructions
3-162
Cop
yr
ight
2000 Dr
aeger
M
edical,
In
c.
sec3_
repa
ir
instr
uc_c.
fm
23.1
Dismounting/Replacing the Patient System
• Unlock patient system
1
by turning screws
2
one quarter of a full rotation (90°)
counterclockwise using an Allen key.
• Lift and remove patient system
1
.
Fig.159:
Removing the patient system
• Place new patient system
1
onto ventilator.
• Secure patient system
1
by turning screws
2
one quarter of a full rotation (90°)
clockwise using an Allen key.
23.2
Safety Valve
In case of repair, replace the complete safety valve.
23.2.1
Testing the Safety Valve
• Check function of safety valve according to Test Certificate.
2
2
1