• Mark silicone hoses for later reassembly and disconnect them from pressure
sensor (flow).
• Remove fixing screws and washers from pressure sensor (flow)
2
.
Fig. 38:
Removing pressure sensor (flow) from Control PCB
• Carefully disconnect the existing pressure sensor (Flow) from terminal strip
1
.
• Replace the old pressure sensor (flow) with a new one.
• Secure new pressure sensor (flow) to Control PCB using fixing screws and
washers
2
.
• Connect silicone hoses to new pressure sensor.
• Mount top cover to control box.
• Mount control box to Gas Box.
• Connect cables and hoses to control box.
• Calibrate pressure sensor (flow) in the standby mode.
• Check pressure sensor (flow) function according to Test Certificate.
D 3
B 3
X13
X23
X16
X4
D7
G 1
D5
L1
L2
+
D
2
2
1
Содержание Fabius 4116029
Страница 2: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 6: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 8: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 68: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 234: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 318: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 324: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 326: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...