Fabius
®
Repair Instructions
3-86
Cop
yr
ight
2000 Dr
aeger
M
edical,
In
c.
sec3_
repa
ir
instr
uc_c.
fm
Before installing the “Pressure Regulator” spare parts set, the pressure regulator
must be closed (depressurized) as follows:
• Unlock adjusting ring
1
(pull in direction of
A
).
• Turn adjusting ring
1
counterclockwise
in direction of
B
as far as it will go.
• Remove nut
2
.
• Mount new “Pressure Regulator” spare parts set (sealing ring, diaphragm, and
locking bolt) to dismounted pressure regulator.
• Mount pressure regulator.
Fig. 71:
Pressure regulator components
Sealing ring
Diaphragm
Locking bolt
A
B
1
2