Fabius
®
Repair Instructions
3-143
Cop
yr
ight
2000 Dr
aeger
M
edical,
In
c.
sec3_r
epai
rinstruc_c.fm
22.1
Pneumatics Control PCB
In case of repair, replace the complete Pneumatics Control PCB.
22.1.1
Testing the Pneumatics Control PCB
• In case of failure, check Pneumatics Control PCB using the debug mode: (no
ventilator error: VEN 0000)
22.1.2
Dismounting/Replacing the Pneumatics Control PCB
• Switch control box off using ON/OFF switch (ON/OFF switch is located on rear
panel of control box).
• Unplug power plug from AC outlet.
• Remove fixing screws
1
from cover plate
2
(rear panel of Fabius) and place cover
plate
2
aside.
Fig.141:
Rear view of Fabius, removing the cover plate
1
1
2