Fabius
®
Repair Instructions
3-20
Cop
yr
ight
2000 Dr
aeger
M
edical,
In
c.
sec3_
repa
ir
instr
uc_c.
fm
51
Ventilation frequency or inspiratory
time out of tolerance.
Motor does not meet
specifications.
Ventilator needs replaced.
Error may also occur if Fabius
is operated while ventilator is
not connected.
81
RAM (Power PCB)
RAM needs replaced.
82
EPROM (Power PCB)
EPROM needs replaced.
83
Watchdog reset (Power PCB)
Software error. Duration
exceeded.