Fabius
®
Repair Instructions
3-148
Cop
yr
ight
2000 Dr
aeger
M
edical,
In
c.
sec3_
repa
ir
instr
uc_c.
fm
• Disconnect connector
1
.
• Mark hoses
2
and
3
for later reassembly and disconnect them from piston pump
connection.
Fig.146:
Top view of the pneumatics, removing the piston pump
• Carefully cut cable tie which fastens electrical cables.
• Push rubber mounts
4
upwards and out through pneumatics plate
5
.
• Mount rubber mounts
4
to new piston pump.
• Mount new piston pump to pneumatics plate
5
by carefully pushing rubber
mounts
4
through pneumatics plate
5
using a screw driver.
Fig.147:
Bottom view of the pneumatics, removing the piston pump
X1
X2
X5
X6
X3
X4
1
2
3
5
4