Fabius
®
Function Description
2-26
Cop
yr
ight
2000 Dr
aeger
M
edical,
In
c.
sec2_
funcde
scr
ip_
b.fm
Manual Ventilation: Expiration
During expiration, the inspiratory valve
1
remains closed thus preventing the
expiratory gas from flowing back into the inspiratory branch.
After releasing the breathing bag
2
, the expiratory gas flows through the open
expiratory valve
3
and the flow sensor
4
into the breathing bag
2
. At the same time,
fresh gas from the fresh-gas port
5
flows through the absorber(s) into the breathing
bag
2
.
Fig. 21:
Manual ventilation (expiration)
3
2
1
4
5