If the message “OK” is displayed, proceed as follows:
• Connect complete breathing system (corrugated hoses, Y-piece, etc.) to circle
absorption system.
• Make sure hose system has no leaks.
• Press “IPPV” key.
• Confirm selection by pressing rotary knob.
• Adjust fresh-gas flow to 0 L/min.
• Adjust PEEP value to 5 hPa, squeeze breathing bag, and check PEEP value
displayed on control box (use standard settings for other PEEP values).
• Adjust PEEP value to 10 hPa, squeeze breathing bag, and check PEEP value
displayed on control box (use standard settings for other PEEP values).
• Check Pmax function using standard settings and adjust Pmax values 25 hPa,
40 hPa, and 70 hPa on control box
(make sure Y-piece is closed)
.
• Measure equivalent leakage current according to Test Certificate.
• Measure protective conductor resistance according to Test Certificate.
• Check function of electronic PEEP valve according to Test Certificate.
Содержание Fabius 4116029
Страница 2: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 6: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 8: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 68: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 234: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 318: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 324: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 326: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...