Fabius
®
Repair Instructions
3-157
Cop
yr
ight
2000 Dr
aeger
M
edical,
In
c.
sec3_r
epai
rinstruc_c.fm
• Hold electronic PEEP valve
1 in place
and remove fixing screws and washers
2
.
Fig.156:
Top view of the pneumatics, replacing the electronic PEEP valve
• Apply Loctite #222 on threads of fixing screws
2
and mount new electronic PEEP
valve to pneumatics.
• Connect hoses to connection ports
3
of electronic PEEP valve (
according to
previously applied markings
).
• Connect connector
4
of electronic PEEP valve to connection
5
on
Pneumatics Control PCB.
• Fasten all electrical cables together using a cable tie.
• Cut off any protruding length of cable tie.
• Mount pneumatics to Fabius using the reverse method of that used for removal.
• Connect gas supplies and electrical connections to Fabius.
• Connect complete breathing system (breathing hoses, Y-piece, breathing bag
etc.) to circle absorption system.
X1
X2
X5
X6
X3
X4
2
2
1
5
4
3
Содержание Fabius 4116029
Страница 2: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 6: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 8: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 68: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 234: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 318: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 324: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 326: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...