4-19
17.1
Test set-up circle system:
COSY leak test:
Switch pressure limitation valve to "MAN"
and set to 70 mbar. Build up test pressure
of
p = 40 mbar.
Test value: Gas flow
*
≤
0.1 L/min.
seal
seal
SEAL
WITH CAP
O2 SENSOR
CONNECTED
SEAL WITH CAP
SV00433
SETTING MAN 70mb
O