Fabius
®
Function Description
2-28
Cop
yr
ight
2000 Dr
aeger
M
edical,
In
c.
sec2_
funcde
scr
ip_
b.fm
Spontaneous Breathing: Expiration
During expiration, the inspiratory valve
1
remains closed thus preventing the
expiratory gas from flowing back into the inspiratory branch.
The APL valve
2
is open, irrespective of the current pressure setting.
The expiratory gas flows through the expiratory hose
3
, the open expiratory valve
4
,
the flow sensor
5
, into the breathing bag
6
. At the same time, fresh gas from the
fresh-gas port
7
flows through the absorber(s)
8
into the breathing bag
6
.
When the breathing bag
6
is full, any excess gas mixture flows through the non-return
valve
9
into the anesthetic gas scavenging system
10
.
Fig. 23:
Spontaneous breathing (expiration)
5
7
3
6
9
10
8
1
2
4
Содержание Fabius 4116029
Страница 2: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 6: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 8: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 68: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 234: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 318: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 324: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 326: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...