General Information About the Flowmeter Block
The flowmeter block is available in several versions. Depending on the drive gases
used, it comes with four flowmeter tubes (2 x N
2
0, 2 x O
2
) or five flowmeter tubes (2 x
N
2
O, 2 x O
2
, and 1 x AIR).
In case of repair, replace (as applicable) the shield, the flow control valves, the
pressure gauges, the rotary knobs, the rotary knob caps, the flowmeter tubes, or the
complete flowmeter block.
18.1
Flowmeter Block Components
Fig. 87:
Flowmeter block (3-gas version)
18.1.1
Testing the Flowmeter Block
• Check the flowmeter block using a separate flowmeter.
S-ORC
Oxygen-Ratio-Controller
O
2
N
2
O
AIR
N2O flowmeter
N2O pressure
N2O control valve
O2 control valve
O2 pressure
O2 flowmeter
Shield
AIR pressure
AIR flowmeter tube
AIR control valve
Содержание Fabius 4116029
Страница 2: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 6: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 8: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 68: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 234: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 318: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 324: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 326: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...