Fabius
®
Repair Instructions
3-40
Cop
yr
ight
2000 Dr
aeger
M
edical,
In
c.
sec3_
repa
ir
instr
uc_c.
fm
14.7
Plug Connector X20 (free)
14.8
Rechargeable Battery Connector X11
The following voltages should be measured according to the current operating mode
(battery-powered, AC-powered, or ON/OFF switch set to OFF): on-load voltage,
charging voltage, off-load voltage.
14.9
Voltage Supply to 5V Voltage Regulator X8
This connector creates the connection between the +24 V voltage supply and the 5V
voltage regulator and motor.
14.10
Status Signals X2
Table 4: Voltage Regulator
Pin
Designation
Value
1
GND
2
FB
approx. +2 V during AC
supply
3
+28V_1
+28 V
4
+28V_2
+28 V
5
FB
approx. +2 V during AC
supply
6
GND
Table 5: Status Signals
Pin
Designation
Value
1
2
Netz_S1
OFF = 8.7 V
ON = 0 V to 3
3
Netz_S2
4
NETZ/BATT
NETZ (AC) = 4.8 V
BATT = 0.6 V to
GND
5
NMI/POWERFAIL
6
not used
Содержание Fabius 4116029
Страница 2: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 6: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 8: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 68: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 234: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 318: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 324: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 326: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...