19.3
Dismounting/Replacing the S-ORC
• Switch control box off using ON/OFF switch (ON/OFF switch is located on rear
panel of control box).
• Unplug power plug from AC outlet.
• Disconnect device from gas supplies.
• Remove fixing screws
1
from flowmeter block
2
.
• Make sure shield
3
does not fall down, and pull flowmeter block
2
slightly out.
Fig.105:
Removing the flowmeter block
• Slide shield
3
of flowmeter block
2
sideways and place it aside.
Fabius
S-ORC
Oxygen-Ratio-Controller
1
2
1
3