• Remove screw and washer
1
from the PEEP valve (see Figure below).
• Disconnect cable connectors
2
from Pneumatics Control PCB.
• Push disconnected cable connectors back into ventilator compartment.
Fig.111:
View of the pneumatics, removing the cable connectors
• Disconnect hose
3
(transparent) and hose
4
(blue) and push them back into the
ventilator compartment.
Fig.112:
Top view of the pneumatics, disconnecting the hoses
X1
X2
X5
X6
X3
X4
2
1
X1
X2
X5
X6
X3
X4
4
3