4-14
8.2
Pressure regulator N
2
O measuring tubes
Supply pressure (5...5,5bar)
Test value: P
dyn
2,0 + 0,2bar
8.3
Pressure regulator Air measuring tubes
Supply pressure (5...5,5bar)
Test value: P
dyn
1,5 + 0,2bar
9.
O2 bypass (flush valve)
Functional test
Test set-up:
Connect mixed-gas outlet to 120 L
flowmeter (7900718)
Test value:
At 5 bar dyn. supply pressure
*
= 39 - 70 L/min.
Test value:
At 3 bar dyn. supply pressure
*
= 27 - 43 L/min.
Pushbutton must return to rest position on
being released.
10.
Ejector outlet
10.1
Sampling coupling (O2)
O2 drive hose
10.2
Capacity
Supply via O2 pressure reducer inlet or
O2-CS inlet.
Air-CS supply depressurised.
Test value:
Sampling coupling
*
> 50 L/min.
Connect coupling to secretion aspirator.
Open shutoff valve of secretion aspirator.
Test objective:
Ejector-type secretion aspirator runs.
11.
Secretion aspirator (optional)
11.1
Ejector-type secretion aspirator assembly
11.1.1
Connecting hose
O L
O L
O
O
O
O L
O L
O
O
C
C L
Содержание Fabius 4116029
Страница 2: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 6: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 8: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 68: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 234: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 318: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 324: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 326: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...