4-28
23.3
Air-CS (if fitted)
Establish O2, N2O and air supply
Pressure gauge for air indicates CS
pressure.
Fully open N2O control valve
Set air flowmeter to
*
= 1 L/min.
Test objective:
Air flowmeter indicates
*
= 1 L/min.
O2 flowmeter indicates
*
= 0.8 L/min.
N2O flowmeter indicates
*
= 1.6 - 2.6 L/
min.
Uncouple O2-CS connector, trigger O2
flush.
Air flowmeter indicates
*
= 1 L/min.
O2 flowmeter indicates
*
= 0 L/min.
N2O flowmeter indicates
*
= 0 L/min.
Uncouple air-CS connector.
Air-CS pressure display falls.
Air flowmeter indicates
*
= 0 L/min.
Close fine control valves, connect CS
hoses.
24.
Functional test of ventilation mode
Switch on Fabius and run through self-
test.
Unit switches to standby on completion of
self-test.
No error messages.
Connect up test thorax 7901082.
24.1
Manual ventilation
Select Man/Spont
Fresh-gas flow
*
= 3 L/min.
APL valve set to MAN, 30 mbar.
Use manual breathing bag to perform
manual ventilation.
24.2
Spontaneous breathing
Test set-up as above:
APL valve set to SPONT/IPPV
Test objective:
Spontaneous breathing must be possible
with test thorax.
O
O
O
O
O
O
O
O
O
O
O
Содержание Fabius 4116029
Страница 2: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 6: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 8: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 68: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 234: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 318: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 324: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 326: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...