Flow Measurement
A moving screen (moving vane) measures the expiratory flow of the patient. A
flow-dependent pressure builds up at the moving vane. This pressure is picked up at
the differential pressure sensor (
∆
P). The differential pressure sensor converts the
pressure into an electronic signal and transmits this signal to the microcontroller.
Fig. 41:
Flow measurement principle
V
.
P
P
∆
P results from the differential pressure
(back-pressure from measuring lines 1 and 2)
moving vane
flow sensor
measuring line
measuring line