14.11
Control PCB
14.11.1 General Information About the Control PCB
In case of malfunctioning, replace the complete Control PCB.
After replacing the Control PCB, carry out "Init System" and calibrate the pressure
sensor, the flow sensor, and the PEEP valve.
• Replace the rechargeable battery at 3-year intervals.
i
As of the first quarter of 1998, new Control PCBs have been
equipped with a Supercap capacitor as standard backup for alarm
sound generation. In this case it is not necessary to remove the
rechargeable battery from the old Control PCB and to install it on the
new one.
Resistor R40 has been set at the factory. Do not change
settings of resistor R40.