• Place piston
1
on the motor spindle plate and secure piston using screws
2
.
Fig.135:
Top view of the motor, mounting the piston
• Work sealing section of rolling seal into its groove.
• Swivel cylinder out of ventilator compartment.
• Seal rolling seal by securing cylinder to bottom of cylinder housing (
do not twist
rolling seal
).
• Secure cylinder to bottom of ventilator using fixing screws.
• Place bag-type rolling seal into cylinder.
1
2