Fabius
®
Repair Instructions
3-17
Cop
yr
ight
2000 Dr
aeger
M
edical,
In
c.
sec3_r
epai
rinstruc_c.fm
0800
Write error.
Ventilation settings (last setting in
the event of a power failure)
Data error.
EEPROM needs replaced.
Data stored incorrectly in the event of a
power failure.
1000
Write error.
Monitoring limit values (last setting
in the event of a power failure)
Data error.
EEPROM needs replaced.
Data stored incorrectly in the event of a
power failure.
2000
Write error.
Monitoring calibration values.
Ventilation calibration values.
Data error.
EEPROM needs replaced.
4000
Error log write error
Data error.
EEPROM needs replaced.
8000
For internal use.
Data error.
EEPROM needs replaced.