Fabius
®
Repair Instructions
3-58
Cop
yr
ight
2000 Dr
aeger
M
edical,
In
c.
sec3_
repa
ir
instr
uc_c.
fm
14.15
Power PCB
In case of failure, replace the complete Power PCB. The EPROM is not included in
the new Power PCB. It must be removed from the old Power PCB and mounted to the
new one.
Fig. 39:
Power PCB, location of EPROM
14.15.1 Testing the Power PCB
• Test Power PCB in the debug mode: (the following is displayed if Power PCB is
OK: VEN 0000, no temperature error).
i
Remove EPROM from Power PCB and mount EPROM onto new
Power PCB.
Make sure to mount EPROM correctly (direction) into EPROM
mount.
D 8
X 9
U 1
D 6
K1
L 1
X 1
X 7
X 3
V4
V5
V18
V12
V11
X 8
EPROM
Jumper plugged = GEC motor
Jumper unplugged = Mavillor motor
Rev. B