Adjusting the Operating Point
• Fully open N
2
O flow control valve on flowmeter block.
• Use O
2
flow control valve to adjust O
2
flow to 0.20 L/min.
The N
2
O flow should be 0.05 L/min (±0.03 L/min).
• If N
2
O flow is not 0.05 L/min (±0.03 L/min), proceed as follows:
• Loosen locknut
1
.
• Turn adjusting screw
2
counter-clockwise to decrease N
2
O flow.
• Turn adjusting screw
2
clockwise to increase N
2
O flow.
Fig.103:
S-ORC, adjusting the operating point
• Secure locknut
1
.
• Re-check operating point.
N
2
O flow should be 0.05 L/min (±0.03 L/min).
N
2
0
0
2
A
E
E
A
2
1