Function Description of the Electronic PEEP Valve
7.3.1
Expiration
The piston pump generates a gas flow. This gas flow is directed to the electronic
PEEP valve.
If an endexpiratory PEEP value has been set on the control box, this value
corresponds to a specific electrical current. The current flows across the coil of the
electronic PEEP valve. The diaphragm closes the crater. The electronic PEEP valve
generates a control pressure. This control pressure is applied to the mechanical
PEEP valve. The patient can only exhale up to the set PEEP value.
7.3.2
Inspiration
During inspiration, the electronic PEEP valve generates a control pressure which
corresponds to the set pressure limit (Pmax).
Fig. 29:
Function diagram of the electronic PEEP valve
electronic PEEP
flow from piston
pump
to atmosphere
to the PEEP control valve
diaphragm
coil
Содержание Fabius 4116029
Страница 2: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 6: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 8: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 68: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 234: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 318: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 324: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 326: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...