Fabius
®
Function Description
2-53
Cop
yr
ight
2000 Dr
aeger
M
edical,
In
c.
sec2_f
uncdescri
p_b
.fm
10.5.9
System Temperature Monitoring
The system temperature sensor is located on the MOSFET output module. It is an
NTC (negative temperature coefficient) thermistor. The microcontroller uses the
system temperature sensor to measure the system temperature.
10.5.10 Quadrature Encoder
The quadrature encoder CF32007NT picks up the number of rotations of the motor
and transmits this data to the microcontroller.
10.5.11 Safety Relay
In the event of a malfunction, the microcontroller switches off the safety relay using
the NOT_AUS (emergency off) signal. As a result, the motor, the electronic PEEP
valve, the piston pump, and the fan can no longer be activated.
10.5.12 Fan Control
If the system temperature inside the control box increases (to approx. 60 °C), the
microcontroller activates the fan.
Содержание Fabius 4116029
Страница 2: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 6: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 8: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 68: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 234: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 318: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 324: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 326: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...