4-18
16.
Low-pressure leak test
Connect mixed-gas socket to pressure
gauge -30...+120 mbar (7901515) or
digital pressure gauge (7901843).
Establish all gas supplies. All flow control
valves closed.
16.1
No build-up in pressure is permissible with
O2 flow control valves closed.
16.2
With Vapor uncoupled, build up pressure
of 120 mbar via O2 flow control valve.
Test value:
Perm. gas flow < 0.1 L/min.
16.3
With Vapor coupled up as for item 16.2
Test value:
Perm. gas flow < 0.1 L/min.
17.
Circle system leak test
Flow sensor connected. Alternatively, flow
for test items 17.1 and 17.2 can be
supplied by anaesthesia machine if:
Machine has previously passed leak test
and result of overall leak test (low-
pressure system of anaesthesia machine
plus circle system) does not exceed
*
≤
0.1 L/min.
Pressure measurement is then performed
with Paw sensor in MAN.
Fabius
Pressure
O
O
O
Содержание Fabius 4116029
Страница 2: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 6: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 8: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 68: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 234: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 318: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 324: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 326: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...