Fabius
®
Repair Instructions
3-55
Cop
yr
ight
2000 Dr
aeger
M
edical,
In
c.
sec3_r
epai
rinstruc_c.fm
• Disconnect silicone hose
1
from pressure sensor (Paw).
• Remove fixing screws and washers
2
from pressure sensor (Paw).
Fig. 36:
Control PCB, removing pressure sensor (Paw)
• Carefully disconnect pressure sensor (Paw) from terminal strip.
• Replace pressure sensor (Paw) with a new one.
• Secure new pressure sensor (Paw) to Control PCB using fixing screws and
washers
2
.
• Connect silicone hose to port
1
of new pressure sensor (Paw) (see Figure
"Control PCB, removing pressure sensor (Paw)").
• Mount top cover to control box.
• Mount control box to Gas Box.
• Connect cables and hoses to control box.
• Calibrate pressure sensor (Paw) in the DrägerService Mode.
• Check pressure sensor (Paw) function according to Test Certificate.
D 3
B 3
X13
X23
X16
X4
D7
G 1
D5
L1
L2
+
D
2
1
2
Содержание Fabius 4116029
Страница 2: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 6: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 8: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 68: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 234: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 318: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 324: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 326: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...