4-27
22.2
N2O supply, cylinder
Open N2O cylinder
(O2 cylinder is open)
Set
*
= 0.8 L/min. at O2 flowmeter
Open all other control valves
N2O flowmeter indicates
*
≈
1.6...2.6 L/
min.
Close N2O cylinder
N2O flowmeter:
Flow drops back to
*
= 0 L/min.
O2 flowmeter:
Flow remains on
*
= 0.8 L/min.
Close O2 and N2O cylinder.
Depressurise system
Close control valves
23.
Testing type of gas, CS operation
All gas supplies depressurised
All pressure gauges for CS depressurised
23.1
O2-CS
Establish O2-CS supply
Pressure gauge for O2 indicates CS
pressure
Actuate O2 bypass valve.
Gas flows
Set O2 flowmeter to
*
= 0.8 L/min.
Fully open N2O control valve.
N2O float at rest.
23.2
N2O-CS
Establish N2O supply
Pressure gauge for N2O indicates CS
pressure
Set O2 flowmeter to
*
= 0.8 L/min.
N2O flowmeter indicates
*
= 1.6...2.6 L/
min.’
Uncouple N2O-CS connector
N2O flowmeter:
Flow drops to
*
= 0 L/min.
O2 flowmeter:
Flow remains on
*
= 0.8 L/min.
O
O
O
O
O
O
O
O
O
O
O
Содержание Fabius 4116029
Страница 2: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 6: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 8: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 68: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 234: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 318: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 324: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 326: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...