• Remove fixing screws
1
from piston and rolling seal
2
.
Fig.120:
Top view of the ventilator, removing the fixing screws
• Remove piston and rolling seal and place them aside.
• Remove screws
3
from motor-cable strain relief
4
.
• Remove screws
5
from light barrier.
• Carefully cut off cable tie.
Fig.121:
Top view of the motor, removing the strain relief
2
1
3
4
5