• Turn Display PCB such that display screen is up.
• Remove label
1
.
• Remove fixing screws
2
.
Fig. 65:
View of dismounted display assembly
• Replace the backlighting lamp with a new one.
• Attach new label below display.
• Mount display to front panel.
• Mount front panel to frame of control box.
• Connect connectors to respective sockets.
• Mount top cover to control box.
• Mount control box to Gas Box.
• Check backlighting lamp function during operation.
HIGH VOLTAGE
C A U T I O N
RISK OF ELECTRIC SHOCK. DISCONNECT THE
ELECTRIC POWER BEFORE SERVICING.