6.2 Explanation
of
Windows
6-13
6
How to Operate Sampling Scope
Status Display
The following statuses of the sampling scope are displayed.
Table 6.2.1-1 Status Display Items
Item
Color
Description
CAL
Orange The correction value of O/E is not
appropriate. Calibrate the O/E module
according to 6.3.2, “Adjusting Dark Current”.
CAL
Yellow There is a difference of ± 2.5 °C or more from
the calibration temperature.
Calibrate the level according to 6.3.1,
CAL
Red
There is a difference of ± 5.0 °C or more from
the calibration temperature.
Calibrate the level according to 6.3.1,
26G CRU Unlock Orange MP2110A-054 Clock Recovery
(Electrical/Optical) is not locked on the input
signal.
53G CRU Unlock Orange MP2110A-055 26G/53Gbaud Clock Recovery
(SM Optical) is not locked on the input
signal.
Free Running
Red
The trigger clock cannot be detected.
PT phase unlock
*
Orange Precision Trigger is not locked on the input
signal.
Trigger setting
wrong
Orange The trigger clock can be detected, but its
frequency does not match the clock frequency
setting value
*: Displayed when the MP2110A-024 is installed.
Содержание BERTWave Series
Страница 26: ...VI...
Страница 74: ...Chapter 1 Outline 1 48...
Страница 166: ...Chapter 4 Screen Operation 4 24...
Страница 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Страница 237: ...6 2 Explanation of Windows 6 43 6 How to Operate Sampling Scope Figure 6 2 4 9 Measure Dialog Box Equalizer Tab...
Страница 239: ...6 2 Explanation of Windows 6 45 6 How to Operate Sampling Scope Figure 6 2 4 10 Measure Dialog Box Mask Test Tab...
Страница 246: ...Chapter 6 How to Operate Sampling Scope 6 52 Figure 6 2 6 2 Time Dialog Box CRU Tab When MP2110A 054 is installed...
Страница 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Страница 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Страница 377: ...8 6 Adding an Option License 8 11 8 Maintenance 10 Click System Menu System Information and confirm the added option...
Страница 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Страница 432: ...Appendix A Specifications A 36...
Страница 457: ...D 3 Sampling Oscilloscope D 11 Appendix D 26G 53G CRU MP2110A 055 Table D 3 7 CRU Output Sensitivity Pass Fail...
Страница 458: ...Appendix D Performance Test Record Form D 12...
Страница 466: ...Index Index 6...