Chapter 7 Performance Test
7-22
6. Click
Scope. Configure the settings as shown below:
Dialog Box
Item
Setting Value
Setup Signal Type
NRZ
Sampling
Mode
Eye
Number
of
Samples
4050
Accumulation
Time Persistency
Time 10.0
sec
Time
Tracking
Symbol Rate: PPG,
Divide Ratio: Sync
Output
Measure -
Display
On
Amplitude/Time Item
(Ch A) Eye Amplitude
7. In the Scope window, click Ch A to measure the amplitude.
8. Adjust the PPG Amplitude so that the Scope amplitude is 50 ±1
mV.
9. Connect the 20 dB fixed attenuator and
ED1 Data In
connector
using a coaxial cable. (Refer to (b) in Figure 7.2.2-1.)
10. Click Start at All Measurements.
11. Click Insert Error at PPG/ED1 once.
12. After measurement, confirm that the error count is 20.
13. Connect coaxial terminators to the
PPG1 Data Out
and
ED1
Data In
connectors. (Refer to (a) in Figure 7.2.2-2.)
14. Attach a 20 dB fixed attenuator to the
PPG1
Data
Out
connector.
15. Connect the 20 dB fixed attenuator and
Ch A In
connector using
a coaxial cable.
16. Repeat steps 7 and 8.
17. Connect the 20 dB fixed attenuator and
ED1
Data
In
connector
using a coaxial cable. (Refer to (b) in Figure 7.2.2-2.)
18. Click PPG/ED Ch1.
19. Set Data Input Condition to Single-Ended XData.
20. Repeat steps 10 to 12.
For ED2 to ED4, perform tests in the same way. For ED3 and ED4 test,
change the Clock Output in step 5 to the following value.
Item
Setting Value
Clock Output
Ch3/4
Содержание BERTWave Series
Страница 26: ...VI...
Страница 74: ...Chapter 1 Outline 1 48...
Страница 166: ...Chapter 4 Screen Operation 4 24...
Страница 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Страница 237: ...6 2 Explanation of Windows 6 43 6 How to Operate Sampling Scope Figure 6 2 4 9 Measure Dialog Box Equalizer Tab...
Страница 239: ...6 2 Explanation of Windows 6 45 6 How to Operate Sampling Scope Figure 6 2 4 10 Measure Dialog Box Mask Test Tab...
Страница 246: ...Chapter 6 How to Operate Sampling Scope 6 52 Figure 6 2 6 2 Time Dialog Box CRU Tab When MP2110A 054 is installed...
Страница 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Страница 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Страница 377: ...8 6 Adding an Option License 8 11 8 Maintenance 10 Click System Menu System Information and confirm the added option...
Страница 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Страница 432: ...Appendix A Specifications A 36...
Страница 457: ...D 3 Sampling Oscilloscope D 11 Appendix D 26G 53G CRU MP2110A 055 Table D 3 7 CRU Output Sensitivity Pass Fail...
Страница 458: ...Appendix D Performance Test Record Form D 12...
Страница 466: ...Index Index 6...