A.3 Sampling Oscilloscope
A-19
A
ppe
ndi
x A
Table A.3.4-3 Jitter Measurement
Item
Specifications
Jitter Measure Setup Settings related to jitter measurement items*
TJ Measurement
BER
Sets the BER to measure TJ(User Define) and Eye Opening with the
measurement result.
Selectable up to 1.0E–001 to 1.0E–018, in 1.0E+001 Step
Fixed RJ
Function to use User-specified value for RJ results
“ON”: Uses the value of Fixed RJ Factor
“OFF”: Uses the calculation result for jitter categorizing.
RJ Value
The value to be used when Fixed RJ is ON
0.01 to 999.99 ps, step 0.01 ps rms
Correction Factor
Function to correct results by Users
“ON”: Corrects the result.
“OFF”: Does not correct the result.
DJ (Scale)
Function to adjust Scale of DJ DJ = DJ (Scale) × Measurement result
0.01 to 999.99, step 0.01
RJ (Scale)
Function to adjust Scale of RJ RJ = RJ (Scale) × Measurement result
0.01 to 999.99, step 0.01
RJ (rms)
Function to adjust rms of RJ
(
)
( )
{
}
2
2
rms
RJ
-
result
t
Measuremen
RJ
=
0.01 to 999.99 ps rms, step 0.01 ps rms
Define Threshold
Definition of Crossing value to measure jitter
Auto:
Automatic adjustment with the use of Crossing value of Scope
Manual:
User input of Crossing to be measured
Manual Crossing
Crossing value to measure jitter
30 to 70%, step 1%
PDJ Measurement Function to switch Pattern Dependent Jitter measurement method.
Available at Advanced Jitter.
“ON”: Displays the result to regard DDJ as PDJ.
“OFF”: Displays the result to regard DDJ as DDJ.
PDJ Standard
Filter
When PDJ Measurement is ON, the following filters are used for PDJ
measurement depending on the Standard. Unit is Hz.
Standard
HP0
HP1
HP1'
HP2
HP'
HP
LP
LP'
STM-0
10
100
–
20 k
–
12 k
400 k
–
STM-1
10
500
–
65 k
–
12 k
1.3 M
500
STM-4
10
1 k
–
250 k
–
12 k
5 M
1 k
STM-16
10
5 k
–
1 M
–
12 k
20 M
5 k
STM-64
10
20 k
10 k
4 M
50 k
12 k
80 M
20 k
STM-256
–
80 k
20 k
16 M
–
–
320 M
–
Measurement Edge
Type
Switching of Edge to measure jitter
Available at Advanced Jitter.
“ALL”:
Measures both rising and falling transition.
“Rising”:
Measures only rising transition.
“Falling”: Measures only falling transition.
Depending on this setting, the result on the graph also displays the
corresponding transition result.
*: Available when MP2110A-096 is installed.
Содержание BERTWave Series
Страница 26: ...VI...
Страница 74: ...Chapter 1 Outline 1 48...
Страница 166: ...Chapter 4 Screen Operation 4 24...
Страница 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Страница 237: ...6 2 Explanation of Windows 6 43 6 How to Operate Sampling Scope Figure 6 2 4 9 Measure Dialog Box Equalizer Tab...
Страница 239: ...6 2 Explanation of Windows 6 45 6 How to Operate Sampling Scope Figure 6 2 4 10 Measure Dialog Box Mask Test Tab...
Страница 246: ...Chapter 6 How to Operate Sampling Scope 6 52 Figure 6 2 6 2 Time Dialog Box CRU Tab When MP2110A 054 is installed...
Страница 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Страница 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Страница 377: ...8 6 Adding an Option License 8 11 8 Maintenance 10 Click System Menu System Information and confirm the added option...
Страница 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Страница 432: ...Appendix A Specifications A 36...
Страница 457: ...D 3 Sampling Oscilloscope D 11 Appendix D 26G 53G CRU MP2110A 055 Table D 3 7 CRU Output Sensitivity Pass Fail...
Страница 458: ...Appendix D Performance Test Record Form D 12...
Страница 466: ...Index Index 6...