Chapter 7 Performance Test
7-24
7.3.1
Required equipment
The following table shows the equipment required for the performance
test.
Before starting performance tests, warm up the MP2110A and measuring
instruments for at least 1 hour.
Table 7.3.1-1 Equipment for Performance Test
Product Name
Required Performance
Recommended
Equipment
Pulse Pattern
Generator or
Signal Generator
Clock frequency:
7 GHz
Amplitude:
0.5
Vp-p
MP2110A
(Anritsu)
DC Power Source*
1
Power voltage:
±2.5 V
Power current:
±50 mA
Setting accuracy:
1% or less
With current limiting function
2400
(Keithley
Instruments)
Light Source*
2
Wavelength:
850, 1310, 1550 nm
Output level range: +3 dBm or more
Level stability:
±0.05 dB
Optical Power
Meter*
2
Wavelength:
750 to 1700 nm
Level range:
–40 to +10 dBm
(100 nW to 10 mW)
Accuracy:
5%
Linearity
±0.05 dB or less
8163B+81623B
(Keysight
Technologies)
Programmable
Optical
Attenuator*
2,
*
4
For single-mode fiber:
Wavelength:
1200 to 1600 nm
Insertion loss:
3 dB or less
Attenuation:
0 to 30 dB
Resolution:
0.1 dB or less
For multi-mode fiber:
Wavelength:
800 to 900 nm
Insertion loss:
3 dB or less
Attenuation:
0 to 30 dB
Resolution:
0.1 dB or less
G0350F
(Anritsu)
G0351F
(Anritsu)
Fixed Attenuator*
3
20 dB K connector
Bandwidth:
DC to 40 GHz
41KC-20
(Anritsu)
Light Source*
4
Wavelength:
1310
nm
Optical fiber:
Single mode
Optical Modulation Amplitude (OMA):
0.64 mW (–2 dBm) or more
Bit rate:
25.78125 Gbit/s
MT1000A,
MU100011A,
G0389A
(Anritsu)
Sampling
oscilloscope*
4
Optical interface
Bandwidth: 30 GHz or more
MP2110A
(Anritsu)
*1: Required for MP2110A-021, MP2110A-023, MP2110A-033, and
MP2110A-043
*2: Required for MP2110A-022, MP2110A-023, MP2110A-025,
MP2110A-026, MP2110A-030, MP2110A-032, MP2110A-033,
Содержание BERTWave Series
Страница 26: ...VI...
Страница 74: ...Chapter 1 Outline 1 48...
Страница 166: ...Chapter 4 Screen Operation 4 24...
Страница 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Страница 237: ...6 2 Explanation of Windows 6 43 6 How to Operate Sampling Scope Figure 6 2 4 9 Measure Dialog Box Equalizer Tab...
Страница 239: ...6 2 Explanation of Windows 6 45 6 How to Operate Sampling Scope Figure 6 2 4 10 Measure Dialog Box Mask Test Tab...
Страница 246: ...Chapter 6 How to Operate Sampling Scope 6 52 Figure 6 2 6 2 Time Dialog Box CRU Tab When MP2110A 054 is installed...
Страница 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Страница 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Страница 377: ...8 6 Adding an Option License 8 11 8 Maintenance 10 Click System Menu System Information and confirm the added option...
Страница 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Страница 432: ...Appendix A Specifications A 36...
Страница 457: ...D 3 Sampling Oscilloscope D 11 Appendix D 26G 53G CRU MP2110A 055 Table D 3 7 CRU Output Sensitivity Pass Fail...
Страница 458: ...Appendix D Performance Test Record Form D 12...
Страница 466: ...Index Index 6...