5.2 PPG/ED
Panel
5-5
5
How to Op
erate BERT
The following table describes the panel.
Table 5.2-1 Setting Items of PPG/ED
Item
Description
Setup/Result
Switches the PPG/ED panel display.
PPG Data/XData
Sets the PPG data signal output On/Off.
XData means
Data
on the front panel.
The right side of the button is lit in green while the data is outputting to
the connector.
Bit Rate*
1
When Reference CLK is Internal:
Select the communication standards by clicking the button.
Set the bit rate when Variable is selected.
Set the offset in the range from –100 to 100 ppm for the set bit rate.
When Reference CLK is Ext Clk In:
Input the external clock and check that the indicator of the Reference
CLK status display is changed to yellow, then click Apply.
When the internal pattern generation circuit is synchronized with
the external clock, the external clock frequency and bit rate are
displayed.
Reference CLK Status
Display
Displays the reference clock synchronization status.
Red:
Reference clock cannot be detected.
Yellow:
Reference clock was detected, but is not synchronized
with pattern data.
Green:
Reference clock is synchronized with pattern data.
PPG Amplitude
Sets the amplitude voltage of the signal output to the
Data Out
connector and
Data
Out
connector at the text box on the right.
When the attenuators are connected to the
Data Out
connector and
Data
Out
connector, set the attenuation at Ext ATT. The amplitude voltage of
the signal output from the attenuator is displayed.
Note:
Connect the attenuators that are the same attenuation to the
Data
Out
connector and
Data
Out
connector.
ED Input Condition
Selects the input method of ED by clicking the button.
Set the threshold voltage of the
Data In
connector and
Data
In
connector
at the text box on the right.
When the attenuators are connected to the
Data In
connector and
Data
In
connector, set the attenuation at Ext ATT.
Note:
Connect the attenuators that are the same attenuation to the
Data
In
connector and
Data
In
connector.
*1: This is set to the following channels as common:
In MP2110A-012: Ch 1 and Ch 2
In MP2110A-014: Ch 1 to Ch 4
Содержание BERTWave Series
Страница 26: ...VI...
Страница 74: ...Chapter 1 Outline 1 48...
Страница 166: ...Chapter 4 Screen Operation 4 24...
Страница 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Страница 237: ...6 2 Explanation of Windows 6 43 6 How to Operate Sampling Scope Figure 6 2 4 9 Measure Dialog Box Equalizer Tab...
Страница 239: ...6 2 Explanation of Windows 6 45 6 How to Operate Sampling Scope Figure 6 2 4 10 Measure Dialog Box Mask Test Tab...
Страница 246: ...Chapter 6 How to Operate Sampling Scope 6 52 Figure 6 2 6 2 Time Dialog Box CRU Tab When MP2110A 054 is installed...
Страница 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Страница 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Страница 377: ...8 6 Adding an Option License 8 11 8 Maintenance 10 Click System Menu System Information and confirm the added option...
Страница 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Страница 432: ...Appendix A Specifications A 36...
Страница 457: ...D 3 Sampling Oscilloscope D 11 Appendix D 26G 53G CRU MP2110A 055 Table D 3 7 CRU Output Sensitivity Pass Fail...
Страница 458: ...Appendix D Performance Test Record Form D 12...
Страница 466: ...Index Index 6...