1.5 Technical
Terms
1-35
1
Outline
Levels p-p, Levels RMS, Level Skews
When measuring the histogram of each level with PAM4 waveform, the
peak value of its width is Levels p-p and the standard deviation is Levels
RMS.
If Sample Timing is Independent, Levels p-p, Levels and RMS are
measured at the following positions.
Level 3
The center of Upper Eye
Level 2
Halfway between the center of Upper Eye and the center
of Middle Eye.
Level 1
Halfway of the center of Middle Eye and the center of
Lower Eye
Level 0
The center of Lower Eye
Level Skews are differences between the positions measured at each level
and their averages.
When Sample Timing is Track to Middle Eye Timing , Levels p-p and
Levels RMS are measured at the center of Center Eye. And Level Skews
are 0.
Figure 1.5-19 PAM4 Levels p-p and Levels RMS
(When Sample Timing is Independent)
Level 2
Level 1
Level 3
Level 0
Levels RMS
Levels p-p
Average of four measured positions
Level Skews
Upper Eye
Center
Lower Eye
Center
Middle Eye
Center
Содержание BERTWave Series
Страница 26: ...VI...
Страница 74: ...Chapter 1 Outline 1 48...
Страница 166: ...Chapter 4 Screen Operation 4 24...
Страница 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Страница 237: ...6 2 Explanation of Windows 6 43 6 How to Operate Sampling Scope Figure 6 2 4 9 Measure Dialog Box Equalizer Tab...
Страница 239: ...6 2 Explanation of Windows 6 45 6 How to Operate Sampling Scope Figure 6 2 4 10 Measure Dialog Box Mask Test Tab...
Страница 246: ...Chapter 6 How to Operate Sampling Scope 6 52 Figure 6 2 6 2 Time Dialog Box CRU Tab When MP2110A 054 is installed...
Страница 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Страница 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Страница 377: ...8 6 Adding an Option License 8 11 8 Maintenance 10 Click System Menu System Information and confirm the added option...
Страница 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Страница 432: ...Appendix A Specifications A 36...
Страница 457: ...D 3 Sampling Oscilloscope D 11 Appendix D 26G 53G CRU MP2110A 055 Table D 3 7 CRU Output Sensitivity Pass Fail...
Страница 458: ...Appendix D Performance Test Record Form D 12...
Страница 466: ...Index Index 6...