7.3 Performance
Test
for
Sampling Oscilloscope
7-27
7
Performance T
est
Figure 7.3.2-3 Amplitude Accuracy Test Connection Diagram
(When using a signal generator)
(3) Procedure
1. In case of Figure 7.3.2-2, connect the
Sync Out
connector and
the
Trigger Clk In
connector using a coaxial cable.
In case of Figure 7.3.2-3, connect the
Trigger Clk In
connector
and the signal generator
output
connector using a coaxial cable.
2. In case of Figure 7.3.2-2, click PPG/ED Ch1. Configure the
settings as shown below:
Item
Setting Value
Reference Clock
Internal
Bit Rate
Variable, 28 Gbit/s, 0 ppm
Sync Output
PPG_1/8 Clk
In case of Figure 7.3.2-3, set the signal generator as follows:
Frequency: 3.5
GHz
Amplitude: 0.5 Vp-p (at sine-wave –2.0 dBm)
Coaxial Cable
+
-
DC Power Supply
Содержание BERTWave Series
Страница 26: ...VI...
Страница 74: ...Chapter 1 Outline 1 48...
Страница 166: ...Chapter 4 Screen Operation 4 24...
Страница 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Страница 237: ...6 2 Explanation of Windows 6 43 6 How to Operate Sampling Scope Figure 6 2 4 9 Measure Dialog Box Equalizer Tab...
Страница 239: ...6 2 Explanation of Windows 6 45 6 How to Operate Sampling Scope Figure 6 2 4 10 Measure Dialog Box Mask Test Tab...
Страница 246: ...Chapter 6 How to Operate Sampling Scope 6 52 Figure 6 2 6 2 Time Dialog Box CRU Tab When MP2110A 054 is installed...
Страница 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Страница 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Страница 377: ...8 6 Adding an Option License 8 11 8 Maintenance 10 Click System Menu System Information and confirm the added option...
Страница 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Страница 432: ...Appendix A Specifications A 36...
Страница 457: ...D 3 Sampling Oscilloscope D 11 Appendix D 26G 53G CRU MP2110A 055 Table D 3 7 CRU Output Sensitivity Pass Fail...
Страница 458: ...Appendix D Performance Test Record Form D 12...
Страница 466: ...Index Index 6...