Chapter 7 Performance Test
7-2
7.1 Performance Test for Pulse Pattern Generator
This section describes the performance test items for the pulse pattern
generator.
Operation frequency range
Frequency accuracy
CAUTION
Connect a 50 Ω terminator to unused input/output
connectors.
7.1.1
Required equipment
The following table shows the equipment required for the performance
test.
Before starting performance tests, warm up the MP2110A and the
measuring instruments for at least 1 hour.
Table 7.1.1-1 Equipment for Performance Test
Product Name
Required Performance
Recommended
Equipment
Sampling
oscilloscope
Electrical interface
Bandwidth: 40 GHz or more
MP2110A
(Anritsu)
Frequency counter
Frequency range:
500 MHz to 20 GHz
Accuracy:
0.1 ppm max.
MF2412C
(Anritsu)
Содержание BERTWave Series
Страница 26: ...VI...
Страница 74: ...Chapter 1 Outline 1 48...
Страница 166: ...Chapter 4 Screen Operation 4 24...
Страница 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Страница 237: ...6 2 Explanation of Windows 6 43 6 How to Operate Sampling Scope Figure 6 2 4 9 Measure Dialog Box Equalizer Tab...
Страница 239: ...6 2 Explanation of Windows 6 45 6 How to Operate Sampling Scope Figure 6 2 4 10 Measure Dialog Box Mask Test Tab...
Страница 246: ...Chapter 6 How to Operate Sampling Scope 6 52 Figure 6 2 6 2 Time Dialog Box CRU Tab When MP2110A 054 is installed...
Страница 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Страница 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Страница 377: ...8 6 Adding an Option License 8 11 8 Maintenance 10 Click System Menu System Information and confirm the added option...
Страница 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Страница 432: ...Appendix A Specifications A 36...
Страница 457: ...D 3 Sampling Oscilloscope D 11 Appendix D 26G 53G CRU MP2110A 055 Table D 3 7 CRU Output Sensitivity Pass Fail...
Страница 458: ...Appendix D Performance Test Record Form D 12...
Страница 466: ...Index Index 6...