Chapter 2 Before Use
2-34
Follow the procedure below to check for electrical overstress to the
MP2110A before connecting DUT (Device Under Test).
Checking the ground connection using a tester
1. Connect DUT and the ground terminal of the MP2110A with a
ground wire.
2. Turn on the power to DUT and MP2110A without connecting DUT to
the I/O connectors of MP2110A.
3. Measure the voltage between DUT’s ground structure and ground
terminal of MP2110A with a tester in AC mode, and confirm the
voltage is 0 V.
If the voltage measured in step 3 is not 0 V, there is a possibility that
MP2110A and DUT may not be connected to a common ground. In this
condition, do not connect the I/O connectors of MP2110A and DUT.
Failure to do so may cause damage to MP2110A due to a voltage
exceeding a rated value.
Review the ground wiring by, for example, changing the location of the
DUT’s ground wire connected to MP2110A so that the voltage measured
in step 3 to be 0 V.
Notes:
If it is impossible to connect a ground wire to DUT, follow the
instructions below so that the voltage to be measured in step 3
becomes 0 V.
•
Check that the 3-pin power cords described in 2.3.2 “Connecting
Power Cord” are used for the MP2110A and DUT.
When the 3-pin power cords are used, take the following actions:
•
There may be a disconnection in the ground terminal of the used
power cord. Replace it with a different 3-pin power cord.
•
There is a possibility that the ground terminal inside the AC
outlets to which the 3-pin power cords of MP2110A and DUT are
connected is not connected to ground. Connect the 3-pin power
cords to different AC outlets.
Содержание BERTWave Series
Страница 26: ...VI...
Страница 74: ...Chapter 1 Outline 1 48...
Страница 166: ...Chapter 4 Screen Operation 4 24...
Страница 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Страница 237: ...6 2 Explanation of Windows 6 43 6 How to Operate Sampling Scope Figure 6 2 4 9 Measure Dialog Box Equalizer Tab...
Страница 239: ...6 2 Explanation of Windows 6 45 6 How to Operate Sampling Scope Figure 6 2 4 10 Measure Dialog Box Mask Test Tab...
Страница 246: ...Chapter 6 How to Operate Sampling Scope 6 52 Figure 6 2 6 2 Time Dialog Box CRU Tab When MP2110A 054 is installed...
Страница 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Страница 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Страница 377: ...8 6 Adding an Option License 8 11 8 Maintenance 10 Click System Menu System Information and confirm the added option...
Страница 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Страница 432: ...Appendix A Specifications A 36...
Страница 457: ...D 3 Sampling Oscilloscope D 11 Appendix D 26G 53G CRU MP2110A 055 Table D 3 7 CRU Output Sensitivity Pass Fail...
Страница 458: ...Appendix D Performance Test Record Form D 12...
Страница 466: ...Index Index 6...