7.2 Performance Test for Error Detector
7-15
7
Performance T
est
(3) Procedure
1. Connect
the
Clk Out
connector and the
Trigger Clk In
connector
of the sampling oscilloscope using a coaxial cable.
2. Connect the coaxial terminator to the
ED1
Data
In
connector
and
PPG1
Data
Out
connector.
3. Attach a 20 dB fixed attenuator to the
PPG1 Data Out
connector.
(Refer to (a) in Figure 7.2.2-1.)
4. Connect the 20 dB fixed attenuator and
Ch A In
connector using
a coaxial cable.
5. Click
PPG/ED Ch1. Configure the settings as shown below:
Item
Setting Value
Reference Clock
Internal
Bit Rate
When MP2110A-093 is not installed:
24 300 000 kbit/s, –100 ppm
When MP2110A-093 is installed:
9 500 000 kbit/s, –100 ppm
Clock Output
Ch1/2
PPG Amplitude
0.5
External ATT
0
Test Pattern (PPG)
PRBS 2^31–1, POS
Test Pattern (ED)
PRBS 2^31–1, POS
ED Input Condition Single-Ended Data
Threshold 0
mV
PPG Data/XData
ON
Gating Cycle
Single
Gating Period
45 s
6. Click
Scope. Configure the settings as shown below:
Dialog Box
Item
Setting Value
Setup Signal Type
NRZ
Sampling
Mode
Eye
Number
of
Samples
4050
Accumulation
Time
Persistency
Time 10.0
sec
Time
Tracking
Symbol Rate: PPG,
Divide Ratio: Clock
Output
Measure -
Display
On
Amplitude/Time Item Selection
(Ch A) Eye Amplitude
7. In the Scope window, click Ch A to measure the amplitude.
8. Adjust the PPG1 Amplitude so that the Scope amplitude is 50±1
mV.
Содержание BERTWave Series
Страница 26: ...VI...
Страница 74: ...Chapter 1 Outline 1 48...
Страница 166: ...Chapter 4 Screen Operation 4 24...
Страница 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Страница 237: ...6 2 Explanation of Windows 6 43 6 How to Operate Sampling Scope Figure 6 2 4 9 Measure Dialog Box Equalizer Tab...
Страница 239: ...6 2 Explanation of Windows 6 45 6 How to Operate Sampling Scope Figure 6 2 4 10 Measure Dialog Box Mask Test Tab...
Страница 246: ...Chapter 6 How to Operate Sampling Scope 6 52 Figure 6 2 6 2 Time Dialog Box CRU Tab When MP2110A 054 is installed...
Страница 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Страница 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Страница 377: ...8 6 Adding an Option License 8 11 8 Maintenance 10 Click System Menu System Information and confirm the added option...
Страница 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Страница 432: ...Appendix A Specifications A 36...
Страница 457: ...D 3 Sampling Oscilloscope D 11 Appendix D 26G 53G CRU MP2110A 055 Table D 3 7 CRU Output Sensitivity Pass Fail...
Страница 458: ...Appendix D Performance Test Record Form D 12...
Страница 466: ...Index Index 6...