Chapter 6 How to Operate Sampling Scope
6-46
Table 6.2.4-9 Items of Mask Test Tab
Item
Description
Mask Test
Turns on and off the mask test on the specified channel.
Eye Mask Select
Select the Mask type. Refer to “Table 6.9.2-1 Mask List”.
Mask Margin Test
Click Run to measure Mask Margin and then stop sampling.
Test Method
Continuous: Measures Mask Margin continuously.
One Shot:
Measures Mask Margin when Run is clicked for Mask
Margin Test.
The Mask Margin measurement is performed so that the failed samples
will be under the value specified at Margin Type.
Margin Type
Selects the method to define the number of failed samples allowed in
Mask Test.
Hit Count:
Defined by the number of failed samples.
Hit Ratio:
Defined by the ratio of failed samples to total samples/UI.
Judged as “Pass” when the number of failed samples is under the
specified value. When Hit Count is 1 (default), failed samples are not
allowed.
Mask Margin
Sets Mask Margin.
When Mask Margin is measured, it is updated to the measured value.
Align Method
Set the method of mask position setting.
Zero/One/Crossing: Places the mask at the midpoint between 0 level
and 1 level and at the midpoint between two
waveform crossing points. This alignment is
performed when Update is clicked for Mask
Alignment or when the Limited measurement
ends.
User Defined:
The user sets the position of the mask.
Alignment Marker
It is displayed when Align Method is User Defined.
Display Off, Display On: Switches marker display.
Center:
Moves the mask to the center of the waveform
display area.
X1,
X, Y1,
Y
When Align Method is User Defined, the mask position and width are
set. Refer to “Figure 6.9.2-3 Example of Adjusting Mask Position
Mask Alignment
It is displayed when Align Method is Zero/One/Crossing.
Click Update to update the mask position.
Mask Area Restriction When setting to On, Angle and Width can be set and the mask area can
be restricted. Refer to “Figure 6.9.2-4 Example of Limiting Mask
When setting Sampling Mode to Advanced Jitter in the Setup dialog box,
Equalizer tab and Mask Test tab cannot be operated.
Содержание BERTWave Series
Страница 26: ...VI...
Страница 74: ...Chapter 1 Outline 1 48...
Страница 166: ...Chapter 4 Screen Operation 4 24...
Страница 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Страница 237: ...6 2 Explanation of Windows 6 43 6 How to Operate Sampling Scope Figure 6 2 4 9 Measure Dialog Box Equalizer Tab...
Страница 239: ...6 2 Explanation of Windows 6 45 6 How to Operate Sampling Scope Figure 6 2 4 10 Measure Dialog Box Mask Test Tab...
Страница 246: ...Chapter 6 How to Operate Sampling Scope 6 52 Figure 6 2 6 2 Time Dialog Box CRU Tab When MP2110A 054 is installed...
Страница 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Страница 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Страница 377: ...8 6 Adding an Option License 8 11 8 Maintenance 10 Click System Menu System Information and confirm the added option...
Страница 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Страница 432: ...Appendix A Specifications A 36...
Страница 457: ...D 3 Sampling Oscilloscope D 11 Appendix D 26G 53G CRU MP2110A 055 Table D 3 7 CRU Output Sensitivity Pass Fail...
Страница 458: ...Appendix D Performance Test Record Form D 12...
Страница 466: ...Index Index 6...