
Chapter 6 How to Operate Sampling Scope
6-50
Table 6.2.5-1 Items of Amplitude Dialog Box
Tab
Item
Description
Scal
e Of
fset
Scale Offset
Sets the level scale.
Scale Sets
vertical
scale.
Offset Sets
vertical
offset.
Attenuation
Sets attenuation amount for external attenuation.
Tracking
Off:
Sets the scale for the active channel.
On:
Sets the scale for all channels to the same value.
Channel Math*
1
Off:
Displays waveforms for Channel A and B separately.
On:
Calculates waveforms for Channel A and B, and then
that calculated result is displayed as Channel A.
Define Function*
1
Sets calculation method between channels.
Scale
Sets vertical scale for calculation result between channels.
Offset
Sets vertical offset for calculation result between channels.
O/E
Input Connector
(Wavelength)*
1,
*
2
Select the wavelength of input light from the following:
MMF connector: 850 nm, User
SMF connector: 1310 nm, 1550 nm, User
Conversion Gain*
2
Indicates the conversion ratio of the O/E.
The setting range is from 1 to 9999 (V/W).
Responsivity*
2
Rate at which photodiode converts optical power to current
The setting range is from 0.001 to 65.535.
Calibration*
2
Automatically adjusts the values of Conversion Gain and
Responsivity when Wavelength is User.
Input Power*
2
Setting used when performing calibration of Conversion Gain
and Responsivity.
Filter Selection*
1
When MP2110A-095 is installed, the filters for 400GbE and
64GFC are added. Since these filters use digital signal
processing, they can be used under the following conditions.
Sampling Mode is set to Coherent Eye
Test Pattern is set to other than Variable
Extinction Ratio
Correction
Sets whether to correct the Extinction Ratio measurement.
This correction is applied to the results of Extinction Ratio
and Outer ExR
Correction Factor
Indicates the correction factor of Extinction Ratio.
The setting range is from –9.99 to 9.99%.
Extinction Ratio is corrected using the following formula:
Extinction Ratio (dB) =
–10log
Zero Level
One Level
– Correction Factor
O/E Calibration
Starts module calibration.
*1: When Sampling Mode is set to Advanced Jitter, it can be operated
while stopping measurement.
*2: When 850 nm, 1310 nm, or 1550 nm is selected at Wavelength, the
values calibrated before factory shipment are set at Conversion Gain
and Responsivity.
Содержание BERTWave Series
Страница 26: ...VI...
Страница 74: ...Chapter 1 Outline 1 48...
Страница 166: ...Chapter 4 Screen Operation 4 24...
Страница 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Страница 237: ...6 2 Explanation of Windows 6 43 6 How to Operate Sampling Scope Figure 6 2 4 9 Measure Dialog Box Equalizer Tab...
Страница 239: ...6 2 Explanation of Windows 6 45 6 How to Operate Sampling Scope Figure 6 2 4 10 Measure Dialog Box Mask Test Tab...
Страница 246: ...Chapter 6 How to Operate Sampling Scope 6 52 Figure 6 2 6 2 Time Dialog Box CRU Tab When MP2110A 054 is installed...
Страница 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Страница 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Страница 377: ...8 6 Adding an Option License 8 11 8 Maintenance 10 Click System Menu System Information and confirm the added option...
Страница 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Страница 432: ...Appendix A Specifications A 36...
Страница 457: ...D 3 Sampling Oscilloscope D 11 Appendix D 26G 53G CRU MP2110A 055 Table D 3 7 CRU Output Sensitivity Pass Fail...
Страница 458: ...Appendix D Performance Test Record Form D 12...
Страница 466: ...Index Index 6...