Chapter 7 Performance Test
7-10
4. Click
PPG/ED Ch1. Configure the settings as shown below:
Item
Setting Value
Reference Clock
Internal
Bit Rate
100GbE/4 (25.78125G), 0 ppm
Clock Output
Ch1/2
PPG Amplitude
0.3 Vp-p
External ATT
0 dB
Test Pattern (PPG)
PRBS 2^9–1, POS
PPG Data/XData
ON
5. Click
Scope. Configure the settings as shown below:
Dialog Box
Item
Setting Value
Setup Signal Type
NRZ
Sampling
Mode
Eye
Number
of
Samples
4050
Accumulation
Time
Persistency
Time
10.0
sec
6. Display the X1 marker by clicking Marker of Scope, and move
the X1 marker to the left cross-point.
7. Connect the coaxial terminator to the
PPG1 Data Out
connector.
8. Connect
the
PPG1
Data
Out
connector to the
Ch A In
connector
of the sampling oscilloscope. (refer to Figure 7.1.4-2).
9. Display the X2 marker by clicking Marker of Scope, and move
the X2 marker to the left cross-point.
10. When the time unit is UI, change Unit in the Scale/Offset tab to
Time by clicking Time of Scope.
11. Measure the time lag of two waveforms using the X1 and X2
markers. Note that the polarities of two waveforms are inverted.
Содержание BERTWave Series
Страница 26: ...VI...
Страница 74: ...Chapter 1 Outline 1 48...
Страница 166: ...Chapter 4 Screen Operation 4 24...
Страница 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Страница 237: ...6 2 Explanation of Windows 6 43 6 How to Operate Sampling Scope Figure 6 2 4 9 Measure Dialog Box Equalizer Tab...
Страница 239: ...6 2 Explanation of Windows 6 45 6 How to Operate Sampling Scope Figure 6 2 4 10 Measure Dialog Box Mask Test Tab...
Страница 246: ...Chapter 6 How to Operate Sampling Scope 6 52 Figure 6 2 6 2 Time Dialog Box CRU Tab When MP2110A 054 is installed...
Страница 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Страница 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Страница 377: ...8 6 Adding an Option License 8 11 8 Maintenance 10 Click System Menu System Information and confirm the added option...
Страница 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Страница 432: ...Appendix A Specifications A 36...
Страница 457: ...D 3 Sampling Oscilloscope D 11 Appendix D 26G 53G CRU MP2110A 055 Table D 3 7 CRU Output Sensitivity Pass Fail...
Страница 458: ...Appendix D Performance Test Record Form D 12...
Страница 466: ...Index Index 6...