Chapter 6 How to Operate Sampling Scope
6-108
13. Click the TJ Measurement BER button to specify the BER to
measure the eye aperture with the Bathtub graph.
Set the following items as needed.
Fixed RJ, RJ Value, Correction Factor, DJ (Scale), RJ (Scale),
RJ (rms), Define Threshold, Manual Crossing
14. Click Sampling to display Sampling Run at the button display.
15. Click Graph on the Result window. Click the Graph selection buttons
to display the jitter graph.
Jitter Analysis (Advanced Jitter)
1. Click
Scope.
2. Click
Setup to set Sampling Mode to Advanced Jitter. Set Signal
Type to NRZ if it is displayed.
3. Click the Limit Type button to set the limitation method for data to
be measured. Proceed to step 5 when None is set.
4. Set Time, Waveforms, Sampling, or Patterns depending on the Limit
Type setting.
5. Click
Time CRU.
6. Click
Scale Offset tab.
7. Set
Pattern
Length.
When Tracking is set to On, select a PPG for Master.
When Tracking is set to Off, set the value for Length.
Note:
The larger the Pattern Length value becomes, the longer it
takes to display the measurement results. It takes up to one
minute.
8. Click
Measure.
9. Click
Amplitude/Time tab.
10. To select an item or items, click on the right column.
11. Click Add. The measurement item is displayed on the Result window.
12. Click Jitter of Measure Setup.
13. Click Algorithm tab.
14. When PDJ measurement is executed, click the PDJ measurement
button to set the display to On. Proceed to step 14 when Off is set.
15. Click the Standard button to specify the standard to be applied to the
PDJ measurement.
16. Click the PDJ Filter button to specify the filter to be applied to the
PDJ measurement.
For the combination of standards and filters, refer to “Table
Settable Standards for PDJ measurement and Filter Sets (Unit Hz)”.
Содержание BERTWave Series
Страница 26: ...VI...
Страница 74: ...Chapter 1 Outline 1 48...
Страница 166: ...Chapter 4 Screen Operation 4 24...
Страница 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Страница 237: ...6 2 Explanation of Windows 6 43 6 How to Operate Sampling Scope Figure 6 2 4 9 Measure Dialog Box Equalizer Tab...
Страница 239: ...6 2 Explanation of Windows 6 45 6 How to Operate Sampling Scope Figure 6 2 4 10 Measure Dialog Box Mask Test Tab...
Страница 246: ...Chapter 6 How to Operate Sampling Scope 6 52 Figure 6 2 6 2 Time Dialog Box CRU Tab When MP2110A 054 is installed...
Страница 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Страница 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Страница 377: ...8 6 Adding an Option License 8 11 8 Maintenance 10 Click System Menu System Information and confirm the added option...
Страница 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Страница 432: ...Appendix A Specifications A 36...
Страница 457: ...D 3 Sampling Oscilloscope D 11 Appendix D 26G 53G CRU MP2110A 055 Table D 3 7 CRU Output Sensitivity Pass Fail...
Страница 458: ...Appendix D Performance Test Record Form D 12...
Страница 466: ...Index Index 6...