Appendix A Specifications
A-28
Table A.3.10-2 Optical Channel Performance (Cont’d)
Item
Specifications
Optical Noise
When MP2110A-022, 023, 025, 026, 032, 033, 035, 036, or 039 is installed:
Conditions
When MP2110A-054
is not installed
When MP2110A-054
is installed*
2
SMF Input, 1310 nm,
OTU4 Filter
Typ. 3.4 µWrms,
Max. 4.3 µWrms
Typ. 4.8 µWrms,
Max. 6.1 µWrms
SMF Input, 1310 nm,
No Filter
Typ.5.4 µWrms
Max. 7.5 µWrms
Typ. 7.6 µWrms
Max. 10.6 µWrms
MMF Input, 850 nm,
OTU4 Filter
Typ. 6.7 µWrms
Max. 8.4 µWrms
Typ. 9.5 µWrms
Max. 11.9 µWrms
MMF Input, 850 nm,
No Filter
Typ. 8.1 µWrms
Max. 10.5 µWrms
Typ. 11.4 µWrms
Max. 14.9 µWrms
When MP2110A-030 is installed:
Conditions
When MP2110A-054
is not installed
When MP2110A-054
is installed*
2
SMF Input, 1310 nm,
OTU4 Filter
Typ. 4.8 µWrms,
Max. 6.1 µWrms
Typ. 6.8 µWrms,
Max. 8.6 µWrms
SMF Input, 1310 nm,
No Filter
Typ.7.6 µWrms
Max. 10.6 µWrms
Typ. 10.8 µWrms
Max. 15.0 µWrms
When MP2110A-040 is installed:
Conditions
When MP2110A-054
is not installed
When MP2110A-054
is installed*
2
SMF Input, 1310 nm,
OTU4 Filter
Typ. 5.8 µWrms,
Max. 7.3 µWrms
Typ. 8.2 µWrms,
Max. 10.4 µWrms
SMF Input, 1310 nm,
No Filter
Typ.7.8 µWrms
Max. 10.6 µWrms
Typ. 11.0 µWrms
Max. 15.0 µWrms
When MP2110A-042, 043, 045, 046, or 049 is installed:
Conditions
When MP2110A-054
is not installed
When MP2110A-054
is installed*
2
SMF Input, 1310 nm,
OTU4 Filter
Typ. 4.1 µWrms,
Max. 5.2 µWrms
Typ. 5.8 µWrms,
Max. 7.4 µWrms
SMF Input, 1310 nm,
No Filter
Typ.5.5 µWrms
Max. 7.5 µWrms
Typ. 7.8 µWrms
Max. 10.6 µWrms
MMF Input, 850 nm,
OTU4 Filter
Typ. 7.0 µWrms
Max. 8.9 µWrms
Typ. 9.9 µWrms
Max. 12.6 µWrms
MMF Input, 850 nm,
No Filter
Typ. 8.6 µWrms
Max. 11.1 µWrms
Typ. 12.1 µWrms
Max. 15.7 µWrms
*2: When the MP2110A-054 is installed, the specification values apply
only to Channel B. To other channels, the specification values at
“When MP2110A-054 is not installed” always apply.
Содержание BERTWave Series
Страница 26: ...VI...
Страница 74: ...Chapter 1 Outline 1 48...
Страница 166: ...Chapter 4 Screen Operation 4 24...
Страница 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Страница 237: ...6 2 Explanation of Windows 6 43 6 How to Operate Sampling Scope Figure 6 2 4 9 Measure Dialog Box Equalizer Tab...
Страница 239: ...6 2 Explanation of Windows 6 45 6 How to Operate Sampling Scope Figure 6 2 4 10 Measure Dialog Box Mask Test Tab...
Страница 246: ...Chapter 6 How to Operate Sampling Scope 6 52 Figure 6 2 6 2 Time Dialog Box CRU Tab When MP2110A 054 is installed...
Страница 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Страница 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Страница 377: ...8 6 Adding an Option License 8 11 8 Maintenance 10 Click System Menu System Information and confirm the added option...
Страница 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Страница 432: ...Appendix A Specifications A 36...
Страница 457: ...D 3 Sampling Oscilloscope D 11 Appendix D 26G 53G CRU MP2110A 055 Table D 3 7 CRU Output Sensitivity Pass Fail...
Страница 458: ...Appendix D Performance Test Record Form D 12...
Страница 466: ...Index Index 6...